Reagent Setup

Leica PELORIS™ User Manual Rev K © Leica Biosystems Melbourne Pty Ltd 2011

96

Drain

1. Connect the wax waste hose and place the end in a suitable container (see 5.4.2 Remote Fill/

Drain Connections).

2. From the Remote fill/drain screen (Reagents menu, Remote fill/drain), select:

The wax stations to drain

The Wax waste icon

Figure 57. Remote fill/drain setup for draining wax chambers

3. Tap Drain to waste to begin the drain.
4. When prompted, check that the wax waste hose is properly connected and feeding into a

suitable container.
Tap OK to begin the drain.

5. The instrument will now drain the chamber(s).

Monitor the drain progress in the status panel. Wax drains may take up to three minutes.
When the drain completes, the chamber state will be Empty.

6. To stop wax solidifying in the drain hose, promptly remove the hose from the drained wax.

At any time during the drain you can tap the Abandon button to terminate all current and pending

fill/drain operations.

WARNING
The wax leaving the waste wax line will be hot and may cause burns. Make sure the wax drains
to a suitable container and stand clear while it drains.

WARNING
Do not remove the wax drain container or hose until the software indicates that the drain has
completed. A cessation of wax flow is not an indication that the drain procedure is complete.
If no wax drains it is likely that the wax hose is blocked. If you remove a blocked hose before
the drain is abandoned hot wax will spurt from the front of the instrument. Abandon the drain
before removing the hose and melting the wax with hot water.

Select the wax chamber or

chambers to drain

Select Wax waste icon

When wax chamber and Wax

waste icon are selected tap

Drain to waste to start the

drain