6 retort scheduling, 1 delayed end times and initial fills, 2 unavoidable reagent clashes – Leica Biosystems PELORIS_PELORIS II User Manual

Page 56: Otocol again (see 3.6.2 unavoidable, Reagent clashes f, 1 delayed, End times and initial fills)

Advertising
background image

Running Protocols

Leica PELORIS™ User Manual Rev K © Leica Biosystems Melbourne Pty Ltd 2011

56

3.6 Retort Scheduling

The Peloris tissue processor allows you to simultaneously run protocols in both retorts. The

automatic scheduling function attempts to assign reagent stations and start times so that there are

no clashes and it may alter your requested end time by starting the protocol early or by inserting a

delay time (see 3.6.1 Delayed End Times and Initial Fills).

Also, when starting a second protocol you may notice that the reagent stations assigned when a

protocol was loaded change as it begins to run. This occurs when the instrument must alter the

assignment to allow for the first protocol’s current reagent requirements.

It is sometimes not possible to schedule a second protocol. This situation and possible remedies are

discussed in 3.6.2 Unavoidable Reagent Clashes. Also, protocols will sometimes fail if a reagent

station unexpectedly becomes unavailable. 3.6.3 Unavailable Reagents describes the best ways to

avoid this situation.

3.6.1 Delayed End Times and Initial Fills

Protocols do not need to start immediately and it is possible to set a required end time that
necessitates a delay before the actual protocol begins. This delay period can extend to many days.

Also, when selecting the ASAP (As Soon As Possible) scheduling option, or if you have requested an

end time that is not achievable, the Peloris tissue processor may be forced to delay the start of the

protocol. During the protocol delay period the Peloris tissue processor will protect your cassettes by

covering them with reagent; this process is called an initial fill.

During the initial fill, the retort is filled with the first scheduled reagent (usually a fixative) to protect

the samples. Unless the reagent is wax no heating or agitation occurs. If the initial step is wax (for

reprocessing or wax only protocols), the retort temperature will be set to wax standby, and the

stirrer will be set to the first step’s speed. Once the initial fill period is over the protocol will run

normally and will finish at the predicted end time.

We recommend that all protocols start with a fixative step (even if only very short) so that a fixative

is used for any initial fill. If there is no fixative step, an initial fill may leave your tissue covered with

dehydrant for a long period and this can cause the tissue to become hard and brittle.

3.6.2 Unavoidable Reagent Clashes

Unavoidable clashes often occur when there are insufficient reagent stations available for both

protocols to satisfy the reagent selection rules (see 4.1.2 Reagent Selection Method). This occurs

more frequently when using “by type” and “by station” protocols as they have limited station

assignment flexibility.

Always ensure sufficient stations of the first reagent group or type so a station is available for an

initial fill.

Advertising