Reagent fill levels in 6.2.1 in, Time and date, Instrument shutdown – Leica Biosystems PELORIS_PELORIS II User Manual

Page 104: Reagent fill levels, Manual drain drip time

Advertising
background image

Settings & Ancillary Operations

Leica PELORIS™ User Manual Rev K © Leica Biosystems Melbourne Pty Ltd 2011

104

The inspection interval for the external system

The maximum value (1000) to limit the number of unwanted warnings

Overridden (contact your service representative to arrange this setting)

See 2.6 External Vapor Removal Systems for more information.

Time and Date

Supervisor-level users can change the time on the instrument on the Instrument settings screen.

Any user can configure the three date and time formats (shortened, normal and extended) used in

the system.

To change the system time tap the Set time button then use the Up and Down buttons as

required to set the correct time.

To change time formats tap the Set date/time format button and then the Modify button

for the format you want to change. Select the required format from the list in the Select time

format or Select date format dialog box.

Instrument Shutdown

Tap the Shutdown instrument button on the Instrument settings screen to turn off the

instrument in an orderly manner. This uses the same shutdown routine as the shutdown button on

the Function bar. When the touch-screen goes blank, turn off the red power button.

See 2.1 Switching On and Shutting Down.

Reagent Fill Levels

The reagent fill levels determine the volume used to fill a retort and the reagent volume needed for

a station to have a full state. You configure the instrument to fill the retorts with sufficient reagent

to process either two or three cassette baskets. If the retorts are set for three baskets the bottle fill

level must also be three baskets, to ensure sufficient reagent to fill the retorts.

You must be a supervisor to change fill levels (on the Instrument settings screen).

To set the retort fill level, tap the Retort fill level button and select two or three baskets.

To set the reagent station fill level, tap the Bottle fill level button and select two or three

baskets.

If you increase the fill volume you must visually check each reagent station to ensure it holds

sufficient reagent. A protocol may be abandoned if a station does not hold sufficient reagent.

Manual Drain Drip Time

Supervisors can change the drip time for manual drain operations on the Instrument settings

screen. The drip time is the length of time the system waits for reagents to drip from cassettes and

retort walls before completing the drain operation.

Tap the Manual drain drip time button then enter the required number of seconds.

Advertising