Running Protocols

Leica PELORIS™ User Manual Rev K © Leica Biosystems Melbourne Pty Ltd 2011

45

In general, avoid this situation by always cleaning the retort immediately after a

processing run, when prompted. Do not put fresh cassettes into a retort that is

showing residue in it, as indicated by the icon at right (formalin residue may be

acceptable).

3.3 Status Screen

Use the Status screen to run protocols and monitor protocol progress. The main features of the
screen are shown in Figure 25.

Figure 25. Status screen

The central area of the screen shows the status of the stations and retorts. Panels at either side of

the screen show protocols loaded for the retorts.

3.3.1 Status Area

3.3.2 Protocol Panels

WARNING
Do not load unprocessed tissue samples into a retort prior to running a cleaning protocol.
Formalin in the residue purged to the wax bath at the start of the cleaning run may damage
tissue on subsequent runs.
If you inadvertently load unprocessed samples into a retort prior to running a cleaning protocol,
remove the samples and attempt to load a processing protocol before loading the cleaning
protocol. The purge before the cleaning run will be skipped.

Retort A

Retort B

16 reagent bottles

4 wax chambers

Retort B protocol panel

Retort A protocol panel

Select/Unload

protocol:

button to load and

unload protocols for

retort A

Run/Pause

button for

retort A

Run/Pause

button for

retort B

Select/Unload

protocol:

button to load and

unload protocols for

retort B