Protocol setup, 1 protocol overview, Protocol overview – Leica Biosystems PELORIS_PELORIS II User Manual

Page 58: Chapter 4, protocol setup : descriptions of pr

Advertising
background image

Leica PELORIS™ User Manual Rev K © Leica Biosystems Melbourne Pty Ltd 2011

58

4

Protocol Setup

Protocols control all aspects of the tissue processing runs carried out on the Peloris. This chapter gives an

overview of Peloris protocols and how to create, edit and view protocols with the software interface.

4.1 Protocol Overview

4.2 Creating, Editing and Viewing Protocols

4.1 Protocol Overview

Protocols consist of sequences of steps applied to tissue in a retort. In each step (with one

exception) the retort is filled with reagent and held for a time under conditions controlled by the

protocol. Protocol steps have the following properties:

the reagent used

the time the reagent is in the retort

the temperature of the reagent

the retort pressure: ambient, pressurized or vacuum, or cycling between pressurized and

vacuum

the retort stirrer speed

the drip time: the time allowed for reagent to drip from the cassettes and retort walls before

ending a drain. Longer drip times reduce carryover.

The one step type that is an exception is the final “dry” step of a cleaning protocol, where no

reagent is used.

There are five protocol types. A protocol’s type sets limits on which reagents can be used and the

order of their use.