8.4 Protocol Step Temperatures

The Peloris tissue processor uses five protocol types to accommodate different processing

functions. Each protocol type has a set of reagent compatibility sequences that comply with the

intended use (see 8.5 Reagent Compatibility Tables for more information). The allowable

temperature range for each step is also dependent on the protocol type. The following sections list

the protocol temperature ranges and typical protocol sequences.

The current wax melting temperature is shown on the Reagent management screen.

8.5 Reagent Compatibility Tables

The reagent compatibility tables determine the allowable reagent sequences. The sequences vary

depending on the operation or protocol type.

Manual Operations

Protocol type

Retort temp. range

for reagent

Retort temp. range for

wax

Wax bath temp.

range

Standard

35 ºC to 65 ºC

2 ºC above wax melt temp.

to 77 ºC

55 ºC to 85 ºC

Xylene-free

35 ºC to 65 ºC

2 ºC above wax melt temp.

to 85 ºC

55 ºC to 85 ºC

Cleaning

35 ºC to 85 ºC

N/A

55 ºC to 85 ºC

Current step

Previous step

= compatible

None

Fixat

iv

es

Deh

ydr

ant

s

Clear

ers

Wa

x

Cleaning solv

ent

s

Cleaning alcohols

Fixatives

Dehydrants

Clearers

Wax

Cleaning solvents

Cleaning alcohols