Reagent Setup

Leica PELORIS™ User Manual Rev K © Leica Biosystems Melbourne Pty Ltd 2011

82

Figure 45. Reagent types screen, temperature thresholds view, using recommended reagent types

for xylene-free processing

Supervisors can edit the default concentration, purity thresholds, and temperature thresholds for all

active reagent types.

You edit a type’s attributes by tapping the appropriate table cell then entering the required value

using the resultant keypad. The attributes update immediately and are applied to all reagent

stations and protocols that use the reagent type. The changes will not affect running protocols.

Lowering temperature thresholds may make protocol steps invalid. You must lower the step

temperature to comply with the new reagent threshold before you can load or run the protocol.

Leica Microsystems does not recommend you lower reagent concentration thresholds. Raising

thresholds can help to correct poor quality processing due to impure reagents.

If you no longer use a reagent in the active list you can remove it to the dormant list, to keep the

active list as small and easy to manage as possible. Select the reagent type by tapping in its name

cell, then tap Remove reagent.

List of active reagent

types

Purity thresholds

and Temperature

thresholds buttons to

switch views

Selected reagent type

Add reagent button:

bring dormant reagent

onto the active list

Remove reagent

button: remove

selected reagent from

the active to the

dormant list

Max. temperatures:

tap cells to change

Defaults/Concentration:

tap cells to change