The pre-defined protocols are:

Factory 1hr Xylene Standard (see Xylene 1 Hour)

Factory 2hr Xylene Standard (see Xylene 2 Hour)

Factory 4hr Xylene Standard (see Xylene 4 Hour)

Factory 8hr Xylene Standard (see Xylene 8 Hour)

Factory 12hr Xylene Standard (see Xylene 12 Hour)

Factory 1 hr Xylene Free (see Xylene-free 1 Hour)

Factory 2 hr Xylene Free (see Xylene-free 2 Hour)

Factory 4 hr Xylene Free (see Xylene-free 4 Hour)

Factory 8 hr Xylene Free (see Xylene-free 8 Hour)

Factory 12 hr Xylene Free (see Xylene-free 12 Hour)

Quick Clean (see Quick Clean)

8.2.3 Xylene Protocols

All the xylene protocols below are designed for use with the xylene bottle configuration in 8.3

Station Configurations.

Xylene 1 Hour

Pre-defined protocol “Factory 1hr Xylene Standard“

Carryover setting in pre-defined protocol: 100

Step Reagent type

Reagent

group

Time

(min)

Temp

(°C)

P/V

Stirrer

Drip time

(s)

1

Formalin

Fixatives

1

Ambient

Ambient

Medium

10

2

Ethanol

Dehydrants

1

Ambient

Ambient

Medium

10

3

Ethanol

Dehydrants

1

Ambient

Ambient

Medium

10

4

Ethanol

Dehydrants

1

Ambient

Ambient

Medium

10

5

Ethanol

Dehydrants

1

Ambient

Ambient

Medium

10

6

Ethanol

Dehydrants

1

Ambient

Ambient. Medium

10

7

Ethanol

Dehydrants

18

45

Ambient

Medium

10

8

Xylene

Clearers

1

Ambient

Ambient

Medium

10

9

Xylene

Clearers

1

Ambient

Ambient

Medium

10

10

Xylene

Clearers

14

45

Ambient

Medium

10

11

Paraffin wax

Wax

2

65

Vacuum

Medium

10

12

Paraffin wax

Wax

1

65

Vacuum

Medium

10

13

Paraffin wax

Wax

14

65

Vacuum

Medium

10

Processing time

1:25:00