Leica Biosystems PELORIS_PELORIS II User Manual

Page 93

Advertising
background image

Reagent Setup

Leica PELORIS™ User Manual Rev K © Leica Biosystems Melbourne Pty Ltd 2011

93

12. The instrument will now fill the bottle(s) via the selected retort.

The fill volume is determined by the reagent fill level set in the Instrument settings screen

(see 6.2.1 Instrument Settings).
Monitor the fill progress in the status panel.
When finished the retort will have an Empty state and the bottle will have a Full state.

At any time during the drain or fill you can tap the Abandon button to terminate all current and

pending fill/drain operations.

If you abandon a drain such that both the retort and bottle are left partially full, you must drain the

retort back to the original bottle to continue. To drain the retort, deselect the Remote icon then

tap the Drain retort button.

Remote Fill and Drain Sequences

The following reagent sequences are recommended when filling and draining multiple bottles:

WARNING
Do not open a retort while it is being used for a remote fill or drain operation as the retort could
be pressurized and may contain hot reagent and fumes. Allow the fill or drain to complete or
abandon the process before opening the retort.

Order

Drain sequence

Fill sequence

1

Fixatives

Cleaning solvents

2

Cleaning alcohols

Clearers

3

Dehydrants

Defatting solvents

4

Defatting alcohols

Defatting alcohols

5

Defatting solvents

Cleaning alcohols

6

Cleaning solvents

Dehydrants

7

Clearers

Fixatives