2.2.3 Magnetic Stirrer

Each retort has as magnetic stirrer that stirs the reagent or wax to ensure temperature consistency

and good tissue penetration. The stirrer is driven by a magnetically-coupled external motor. The

stirrer speed can be controlled for each protocol step and is easily removed for cleaning (see Clean

Retorts in 7.2 Daily Tasks).

Figure 6. Magnetic stirrer (1)

2.2.4 Cassette Baskets

There are two Peloris basket types: the configurable high-capacity basket that accommodates the

maximum number and type of cassettes; and the spaced basket that ensures optimal reagent flow.

The high-capacity cassette baskets accept most common cassette types and include configurable

dividers that allow for different cassette sizes and packing densities.

The spaced baskets include dividers that ensure optimum reagent flow with minimal carryover. This

basket type accepts 72 standard cassettes. Spaced baskets must be used for all xylene-free

protocols.

Cassette baskets stack into the retorts with the lids upwards and handles dropped flat. Make sure

the first basket sits flat on the rack on the pins at the bottom of the retort (Figure 8). Stack

additional baskets so they sit flat on the lower ones – crooked baskets can leave cassettes exposed,

damaging the tissue in them.

Figure 7. Placing a basket into a retort

Figure 8. Retort pins