Xylene 8 hour – Leica Biosystems PELORIS_PELORIS II User Manual

Page 129

Advertising
background image

Reference

Leica PELORIS™ User Manual Rev K © Leica Biosystems Melbourne Pty Ltd 2011

129

Xylene 6 Hour

Not a pre-defined protocol

Xylene 8 Hour

Pre-defined protocol “Factory 8hr Xylene Standard“

Carryover setting in pre-defined protocol: 25

Step Reagent type

Reagent

group

Time

(min)

Temp

(°C)

P/V

Stirrer

Drip time

(s)

1

Formalin

Fixatives

15

45

Ambient

Medium

10

2

Ethanol

Dehydrants

15

45

Ambient

Medium

10

3

Ethanol

Dehydrants

15

45

Ambient

Medium

10

4

Ethanol

Dehydrants

15

45

Ambient

Medium

10

5

Ethanol

Dehydrants

15

45

Ambient

Medium

10

6

Ethanol

Dehydrants

30

45

Ambient

Medium

10

7

Ethanol

Dehydrants

45

45

Ambient

Medium

10

8

Xylene

Clearers

20

45

Ambient

Medium

10

9

Xylene

Clearers

20

45

Ambient

Medium

10

10

Xylene

Clearers

45

45

Ambient

Medium

10

11

Paraffin wax

Wax

30

65

Vacuum

Medium

10

12

Paraffin wax

Wax

30

65

Vacuum

Medium

10

13

Paraffin wax

Wax

45

65

Vacuum

Medium

10

Processing time

6:08:00

Step Reagent type

Reagent

group

Time

(min)

Temp

(°C)

P/V

Stirrer

Drip time

(s)

1

Formalin

Fixatives

20

45

Ambient

Medium

10

2

Ethanol

Dehydrants

20

45

Ambient

Medium

10

3

Ethanol

Dehydrants

20

45

Ambient

Medium

10

4

Ethanol

Dehydrants

20

45

Ambient

Medium

10

5

Ethanol

Dehydrants

20

45

Ambient

Medium

10

6

Ethanol

Dehydrants

40

45

Ambient

Medium

10

7

Ethanol

Dehydrants

60

45

Ambient

Medium

10

8

Xylene

Clearers

30

45

Ambient

Medium

10

9

Xylene

Clearers

30

45

Ambient

Medium

10

10

Xylene

Clearers

60

45

Ambient

Medium

10

11

Paraffin wax

Wax

40

65

Vacuum

Medium

10

12

Paraffin wax

Wax

40

65

Vacuum

Medium

10

13

Paraffin wax

Wax

60

65

Vacuum

Medium

10

Processing time

8:08:00

Advertising