4 replacing reagents, 1 remote fill/drain screen, Replacing reagents remote fill/drain screen – Leica Biosystems PELORIS_PELORIS II User Manual

Page 88: 1 remote fill/drain, Screen, Ee 5.4 replacing reagents

Advertising
background image

Reagent Setup

Leica PELORIS™ User Manual Rev K © Leica Biosystems Melbourne Pty Ltd 2011

88

5.4 Replacing Reagents

Always replace reagents as soon as possible after the system alerts you.

There are two methods to replace reagent in bottles:

Remote fill/drain — pump the old reagent out and new reagent back in without removing

the bottle, using commands on the Remote fill/drain screen.

Manually — remove the bottle, drain and refill, then replace on the instrument.

For wax you must drain the chamber using the Remote fill/drain screen, fill the chamber

manually, and then update the software.

See sections below:

5.4.1 Remote Fill/Drain Screen

5.4.2 Remote Fill/Drain Connections

5.4.3 Replacing Reagent – Remote Fill and Drain

5.4.4 Replacing Reagent – Manual

5.4.5 Replacing Wax

5.4.6 Filling and Draining Retorts

5.4.1 Remote Fill/Drain Screen

Use the Remote fill/drain screen (Reagents menu, Remote fill/drain) to drain the wax baths

and to fill and drain reagent bottles without removing them from the instrument. You are able to

fill/drain stations individually or as a group of compatible stations in a single operation. You can also

fill and drain retorts from this screen to enable recovery from partially completed fill/drain

operations.

All users can perform the functions on the Remote fill/drain screen.

Advertising