Environmental

Maximum ambient temperature: ........................................................................................ 35 °C

Minimum ambient temperature: ...........................................................................................5 °C

Humidity (noncondensing): ...................................................................................10 to 80% RH

Altitude: .......................................................................................... 0 to 2000 m above sea level

Sound pressure level output (at 1 m): .............................................................................< 65 dB

Maximum heating energy output: ............................................................1450 W (100 to 120 V~)

2150 W (220 to 240 V~)

Electrical

Operating voltage: ...............................................................................................100 to 120 V~

220 to 240 V~

Operating current (maximum): ....................................................................15 A (100 to 120 V~)

10 A (220 to 240 V~)

Mains frequency: ................................................................................50 –60 Hz (100 to 120 V~)

50 Hz (220 to 240 V~)

Power consumption: ...............................................................................1450 W (100 to 120 V~)

2150 W (220 to 240 V~)