Xylene-free Configuration

This is the suggested station configuration when running xylene-free protocols (see 8.2.4 Xylene-

free Protocols). See note below for special conditions for initial bottle setup.

*All alcohol dehydrants after fixative should be labeled and named in the software “85% Ethanol”. However, if
setting up bottles on a new instrument or replacing all the ethanol bottles at once, the first two bottles should
be filled with 70% ethanol (but still be named “85% Ethanol”). As operation continues and you are prompted
to replace degraded bottles, replace with 85% ethanol. Other, initially higher-concentration bottles will have
degraded, so there will always be at least one bottle at a low concentration.

Station

Reagent type

Bottle 1

Formalin

Bottle 2

Formalin

Bottle 3

85% Ethanol*

Bottle 4

85% Ethanol*

Bottle 5

85% Ethanol

Bottle 6

80/20 Ethanol/IPA

Bottle 7

80/20 Ethanol/IPA

Bottle 8

80/20 Ethanol/IPA

Bottle 9

IPA

Bottle 10

IPA

Bottle 11

IPA

Bottle 12

IPA

Bottle 13

Waxsol™

Bottle 14

Waxsol™

Bottle 15

Cleaning ethanol

Bottle 16

Cleaning ethanol

Wax 1

Wax

Wax 2

Wax

Wax 3

Wax

Wax 4

Wax