Connecting to an external system – Leica Biosystems PELORIS_PELORIS II User Manual

Page 36

Advertising
background image

Hardware

Leica PELORIS™ User Manual Rev K © Leica Biosystems Melbourne Pty Ltd 2011

36

For instructions on changing to an external vapor system, see Connecting to an External

System below.

For instructions on returning to the internal carbon filter system, see Returning to the Internal

Filter System.

Some instruments do not have the vapor select valve fitted. However your service representative

can easily connect these instruments to an external vapor system.

Connecting to an External System

Use the following procedure to connect the instrument to an external vapor system.

1. Connect the instrument’s vapor outlet into the external system (see item 2 in Figure 18).
2. Turn the vapor select valve a quarter turn anticlockwise to direct vapors to the vapor outlet

(see Figure 19).
Note: to access the valve you may need to remove a blanking plug.

Figure 19. Vapor select valve in the external system position

3. Set the carbon filter threshold to one of the following options.

(i) The inspection interval for the external system (see 6.2.1 Instrument Settings).
(ii) The maximum value (1000) to limit the number of unwanted warnings (see 6.2.1

Instrument Settings).

(iii) Overridden (contact your service representative to arrange this setting).

WARNING
When the vapor select valve is in the external position, you must ensure an external vapor
system is correctly installed or potentially dangerous fumes will be released into the laboratory
environment.

Advertising