7 staining artifact, Staining artifact – Leica Biosystems PELORIS_PELORIS II User Manual

Page 147

Advertising
background image

Troubleshooting

Leica PELORIS™ User Manual Rev K © Leica Biosystems Melbourne Pty Ltd 2011

147

9.2.7 Staining Artifact

Is there cellular

detail?

Under/Over processed

Change or

replace solutions

in reagent bottles

Ensure protocol

step times are

adequate

Is there formalin

pigment?

pH of formalin is too

high

Fix tissue in neutral

buffered formalin,

and change

regularly

Is there coarse or

fine cracking

?

Over processed

Change or replace

solutions in reagent

bottles

Ensure protocol step times

are adequate

Is there zonal

fixation?

Under fixed

Make sure tissue is placed in

an adequate amount of

formalin for an adequate

amount of time. As soon as

possible after removal from

patient

Ensure tissue is

not too big for

cassette

Does the tissue have

blue hue or smudge?

Water present in

tissue

Change or replace solutions in

reagent bottles

Increase step time in

dehydration

See sections

9.2.1, 9.2.2, 9.2.3

No

Yes

Yes

Yes

See sections

9.2.1, 9.2.2, 9.2.3

Inadequate

fixation