4 Introduction to the operation

746 VA Trace Analyzer / 747 VA Stand

4-26

2

Switch to dialog page "RESULTS/Evaluation"

•

Press the softkey

<Eval.>

. The dialog page "RESULTS" with the sub-

page "Evaluations" appears showing an overview of the substance and

peak evaluations of the determination loaded in the working storage:

Method : TESTPb RESULTS

Determ.: 03201614 Evaluations

Substance : Pb Comments

Mass conc.: 1.032 g/L Mass : 103.2 ug ––––––––––––––––––

MC.dev. : 0.011 g/L (1.05%) Add.mass : 100 ug

Cal.dev. : - V0.sample: 100 uL

VR U/mV I/nA I.mean Std.dev. I.delta Comments

––– ––––– –––––– –––––– ––––––– ––––––– ––––––––––––––––––

00 -355 -195.9 -194.7 1.569

01 -355 -195.3

02 -355 -192.9

10 -356 -385.2 -385.5 0.4942 -190.8

11 -356 -386.1

12 -356 -385.3

20 -357 -568.1 -568.5 1.968 -182.9

21 -357 -566.7

22 -357 -570.6

Select

subst.

Suspend/

activate

New

calcul.

>>

Calcul.

>>

Calibr.

•

You will find a brief explanation of the result data shown here in the

description of the full report in section 4.2.6.

•

On this subpage you have the possibility to ignore undesired peak

results (e.g. spikes) for subsequent calculations. Proceed as follows:

−

Use the cursor keys to select the desired result by its VR num-

ber.

−

Press the softkey

<Suspend/activate>

. In addition to the VR

number, the marking

!

appears to show that this peak result will

be ignored in subsequent calculations (this marking can also be

canceled using the same softkey).

−

Press the softkey

<New calcul.>

. This initiates a new calculation

whose effects on the substance evaluation are immediately ap-

parent on this page.

3

Switch to dialog page "RESULTS/Calibration"

•

Press the softkey

<Calibr.>

. The dialog page "RESULTS" with the

subpage "Calibration" appears showing the results of the substance cali-

brations: