7.4 Troubleshooting

746 VA Trace Analyzer / 747 VA Stand

7-39

Peak is in the highest

µ

A range

•

The sample volume is too large: reduce the volume. Carry out the analysis

again.

•

The enrichment period (for HMDE) is too long: Reduce the time.

•

If necessary use a SMDE or DME electrode instead of HMDE.

Double peak

•

Check MME. If necessary, change capillary (see section 3.4.8) or replace

sealing needle (see section 3.4.9).

•

Organic components interfere with the analysis? Carry out a UV digestion or

other suitable sample preparation.

•

If a second element is present at the same potential: add this element to the

sample and carry out the analysis again. If the second peak has become

higher then the second element is present.

•

Might it be possible to selectively mask this second element with a complex

former?

•

Electrolyte solution too old: make up a new one. Its working life with organic

additives may be as short as 1 day or less.

•

For Cu: work without chlorides in the electrolyte or increase the chloride con-

centration massively.

•

Try another measurement mode like AC1. If one substance is reversible and

the second one irreversible, only the reversible substance is detected by

AC1.

Standard addition peaks displaced

•

If HMDE is used potential displacements of more than 20

…

30 mV are often

normal and have to be accepted; particularly in adsorption voltammetry.

•

Standard solutions have been made too acidic.

•

Buffering capacity of the electrolyte is not sufficient: increase electrolyte vol-

ume.

No addition

•

Has the correct standard solution been used or is the concentration of the

solution too low: increase the volume of the standard addition or use a higher

concentration or reduce the sample amount accordingly.

•

If organic components are present: carry out a UV digestion or other suitable

sample preparation.

•

Concentration of the analyte is too high: dilute.

•

Addition solution with metal complexing solution (time reaction).