Metrohm 746 VA Trace Analyzer User Manual
Page 251

5.5 Operation
746 VA Trace Analyzer / 747 VA Stand
5-100
Parameter
Range; Default value / Meaning
I.peak
display only
Peak height
Determined peak height in A.
I.peak
is shown for curves recorded with one of the following meas-
urement modes:
DPMODE, SQWMODE, AC1MODE
DCTMODE with SSWEEP (peaks)
CYCMODE with only 1 ramp
I
U
I.peak
I.wave
display only
Wave height
Determined wave height in A.
I.wave
is shown for curves recorded with one of the following meas-
urement mode DCTMODE and SWEEP (for waves).
I.wave
I
U
I.abs, I.rel
display only
Current evaluation quantity for AC2 curves
Determined current evaluation quantity in A.
I.abs
or
I.rel
is shown for curves recorded with the measurement
mode AC2MODE.
I.abs
Type = AC2 abs.
I
f
I
r
I.rel = I
f
+ I
r
I
U
Type = AC2 rel.
I
U
P.peak
display only
Peak area
Determined peak area in W.
P.peak
is shown for curves recorded with one of the following meas-
urement modes:
DPMODE, SQWMODE, AC1MODE
DCTMODE with SSWEEP (peaks)
CYCMODE with only 1 ramp
In an evaluation with
Scope = f.half
or
Scope = r.half
, the area up
to the peak maximum is shown here.