Metrohm 746 VA Trace Analyzer User Manual

Page 510

Advertising
background image

7.3 Error messages

746 VA Trace Analyzer / 747 VA Stand

7-11

Error message

Meaning/Causes

Rectification

Exit

DATA CARD - ERROR 9:
File name not found

Error in copying from or to data

card

−

Attempt copying again

−

If error reappears, delete file in

question

−

If error reappears with this data card,

save all files and reformat data card

E/C

DATA CARD - ERROR 10:
File already closed

Error in copying from or to data

card

−

Attempt copying again

−

If error reappears, delete file in

question

−

If error reappears with this data card,

save all files and reformat data card

E/C

DATA CARD - ERROR 11:
File already opened

Error in copying from or to data

card

−

Attempt copying again

−

If error reappears, delete file in

question

−

If error reappears with this data card,

save all files and reformat data card

E/C

DATA CARD - ERROR 12:
Card is already full

No more free storage space on

data card

Use new data card or clear data

E/C

DATA CARD - ERROR 13:
Invalid mode

Error in copying from or to data

card

−

Attempt copying again

−

If error reappears, delete file in

question

−

If error reappears with this data card,

save all files and reformat data card

E/C

DATA CARD - ERROR 14:
File is protected

The selected file can not be

copied to the data card as a

protected file with the same

name is already present

(overwrite protection)

Disable overwrite protection with

<Secure/Unsecure>

or change file

name

E/C

DATA CARD - ERROR 15:
Unexpected end of file

Error in copying from or to data

card

−

Attempt copying again

−

If error reappears, delete file in

question

−

If error reappears with this data card,

save all files and reformat data card

E/C

DATA CARD - ERROR 16:
Card is write protected

The data card is write protected

Disable overwrite protection or use a

different data card

E/C

DATA CARD - ERROR 17:
Undefined error

Unspecified error in data traffic

with the data card

Press Reset key

E/C

DATA CARD - ERROR 18:
Bad battery charge
detected

Charge of data card battery

bad

−

Copy all data on data card to instru-

ment memory

−

Change battery

−

Recopy data to data card

E/C

DATA PROC - ERROR 1:
Not enough measuring
points

Too few measured points

available for the smoothing

Extend measurement range or reduce

U.step

E/C

DATA PROC - ERROR 2:
Current overload, data
not evaluable

More than 2 overflow measured

values have appeared in

succession, the curve can thus

not be evaluated

Start new measurement or remove

overflow measured values manually

using the spike elimination

E/C

DATA PROC - ERROR 3:
No background sweep
found

No background sweep avail-

able

Rerecord measurement with back-

ground compensation

E/C

DIAGNOSIS - ERROR 1:
No VA Stand found

Diagnostic fault; connection to

747 VA Stand broken:

−

connecting cable not

plugged in

−

wrong address

Check connection to 747 VA Stand:

−

plug in connecting cable

−

set address correctly and restart 746

VA Trace Analyzer

E/C

Advertising