Parameter
Range; Default value / Meaning
dU.rear
display only
Voltage difference to rear base point
Absolute voltage difference determined between the base point at the
peak/wave rear half and the peak voltage
U.peak
or the half-wave
potential
U/2
in mV.
S.front
display only
Slope of baseline at front base point
Slope of the baseline determined at the base point at the peak/wave
front half.
With peaks with a baseline of the
linear
type, this parameter is not
shown.
S.rear
display only
Slope of baseline at rear base point
Slope of the baseline determined at the base point at the peak/wave
front half.
With peaks with a baseline of the
linear
type, this parameter is not
shown.
linear, whole
dU.frontdU.rear
S.rear
S.front
S.rear
S.front
S.front
S.rear
linear, DCT
dU.frontdU.rear
expon., whole
dU.frontdU.rear
polynom, whole
dU.frontdU.rear