5.5 Operation

746 VA Trace Analyzer / 747 VA Stand

5-92

Parameter

Range; Default value / Meaning

Techn.

display only

Calibration technique

Calibration technique selected for the evaluation of the substance
(see sections 5.6.3 and 6.9):

std.add.

Standard addition

cal.crv.

Calibration curve

int.std.

Internal standard

Y.reg/offset

display only

Regression value or intercept (variable KY:<subst>)

Depending on the calibration technique selected, the following is
shown under this parameter (see sections 5.6.3 and 6.9):

std.add.

Current or area value for the sample in A, W or s cal-
culated from the standard addition curve

cal.crv.

Intercept of the calibration curve

int.std.

No display

This value can be used in formulae as

KY:<subst>

(

<subst>

=

substance name) for further calculations. It is put out in such cases in
the base units A, W or s without prefix (see section 5.6.4).

Slope

display only

Slope or proportionality factor (variable KS:<subst>)

Depending on the calibration technique selected, the following is
shown under this parameter (see sections 5.6.3 and 6.9):

std.add.

Slope of the standard addition curve
(in AL/g, WL/g or sL/g)

cal.crv.

Slope of the calibration curve (in AL/g, WL/g or sL/g)

int.std.

Proportionality factor between analyte and reference
substance (in AL/g, WL/g or sL/g)

This value can be used in formulae as

KS:<subst>

(

<subst>

=

substance name) for further calculations. It is put out in such cases in
the base units AL/g, WL/g or sL/g without prefix (see section 5.6.4).

Nonlin.

display only

Non-linearity factor (variable KN:<subst>)

Non-linearity factor of the calibration curve or the standard addition
curve (in [AL/g]

4

, [WL/g]

4

or [sL/g]

4

, see sections 5.6.3 and 6.9).

This value can be used in formulae as

KN:<subst>

(

<subst>

=

substance name) for further calculations. It is put out in such cases in
the base units without prefix (see section 5.6.4).

Mean deviat.

display only

Mean deviation of standard addition or calibration curve

(variable KD:<subst>)

Calculated mean deviation of the measured values about the standard
addition or calibration curve in A, W or s.
The size of

Mean deviat.

can be used as a measure of the quality of

a standard addition or calibration curve.

This value can be used in formulae as

KD:<subst>

(

<subst>

=

substance name) for further calculations. It is put out in such cases in
the base units A, W or s without prefix (see section 5.6.4).

Advertising