5.7 Program instructions
746 VA Trace Analyzer / 747 VA Stand
5-304
Comments
•
The electrode type must be defined before the
TSWEEP
instruction.
Only
HMDE
or
RDE
are admissible.
•
The measurement mode must be defined before the
TSWEEP
instruction. Admissible are
DCTMODE
,
DPMODE
,
AC1MODE
,
AC2MODE
or
SQWMODE
, but not
PSAMODE
or
CYCMODE
.
•
Apart from the
TSWEEP
instruction, no other SWEEP instruction is
allowed in the same segment.
•
If the current measurement should be performed without measured
value recording, the
TSWEEP
instruction must be replaced by the
MEAS
instruction. In contrast to
TSWEEP
, this may be in the same
segment as the measurement sweep.
•
If the measurement voltage
U.meas
should not be switched off at the
end of the program, enter a suitable standby voltage
U.standby
in a
ØMEAS
instruction.
•
Maximum 1000 measured values can be recorded in a segment. The
number of measured values is calculated as follows:
Measurement duration
t/s
Number of measured values = –––––––––––––––––––––––––
Time interval
t.step
see also
HMDE, RDE, MEAS