5.7 Program instructions

746 VA Trace Analyzer / 747 VA Stand

5-304

Comments

•

The electrode type must be defined before the

TSWEEP

instruction.

Only

HMDE

or

RDE

are admissible.

•

The measurement mode must be defined before the

TSWEEP

instruction. Admissible are

DCTMODE

,

DPMODE

,

AC1MODE

,

AC2MODE

or

SQWMODE

, but not

PSAMODE

or

CYCMODE

.

•

Apart from the

TSWEEP

instruction, no other SWEEP instruction is

allowed in the same segment.

•

If the current measurement should be performed without measured
value recording, the

TSWEEP

instruction must be replaced by the

MEAS

instruction. In contrast to

TSWEEP

, this may be in the same

segment as the measurement sweep.

•

If the measurement voltage

U.meas

should not be switched off at the

end of the program, enter a suitable standby voltage

U.standby

in a

ØMEAS

instruction.

•

Maximum 1000 measured values can be recorded in a segment. The
number of measured values is calculated as follows:

Measurement duration

t/s

Number of measured values = –––––––––––––––––––––––––

Time interval

t.step

see also

HMDE, RDE, MEAS