Metrohm 746 VA Trace Analyzer User Manual

Page 246

Advertising
background image

5.5.3 CURVES

746 VA Trace Analyzer / 747 VA Stand

5-95

Softkeys

Meaning

Display

segment

Display segment

Switching to the display of the segment associated with the selected
substance.

Select

subst.

Select substance curves

The selection of substance and VR numbers determines what sub-
stance curves are displayed. First, the softkeys carrying the substance
name and segment name must be used to select the substance:

Subst.#

Segment#

Selection of the substance

Subst.#

allocated to the segment

Segment#

.

This is followed by entry of the numbers for variation

V

and replication

R

. For this, the system window shows the message:

VR max: ##

Maximum selectable VR values

Enter VR: [*0]

Entry of VR

The following possibilities are available for the entry of VR:

##

Entry of two numbers

##

for the display of a single

curve.

*#

Entry of an asterisk

*

and a number

#

for the display

of all variations with the same replication (shown by

(grouped VR: *#)

in the title line).

#*

Entry of a number

#

and an asterisk

*

for the display

of all replications of the same variation (shown by

(grouped VR: #*)

in the title line).

**

Entry of two asterisks

**

for the display of all curves

(shown by

(grouped VR: **)

in the title line).

Some of the possible entries can be selected directly by pressing the
appropriate softkeys. Without entry of VR, the default value

*0

is

effective for all variations with replication 0 as standard.

Select

curve

Select curve

The following curves can be selected for each substance using the
softkeys:

Original

Display of the original curve (original measured values)

Valid measured values

k

Recognized spikes

¡

Invalid measured values (overflow)

I

U

k

k

k

In determinations with

PSAMODE, the original

curve can not be dis-

played.

12

Advertising