Metrohm 746 VA Trace Analyzer User Manual
Page 554

7.6 Diagnosis
746 VA Trace Analyzer / 747 VA Stand
7-55
•
Attach electrode cable WE to clamping screw WE-D.
•
<ENTER>
2 diagrams are printed out and must be assessed as follows:
Segment WE-L (DCTMODE):
−
In principle, the plotted diagonal must be straight. Small deviations
are partly due to the resolution of the printer.
−
Current at voltage
– 200 mV:
– 1.6
µ
A ... – 2.4
µ
A*
−
Current at voltage
+200 mV:
+1.6
µ
A ... +2.4
µ
A*
* These tolerance values refer to illustrative scatters, that is to the differences
between individual instruments. The value determined for a specific instrument
shows less scatter. The deviation depends mainly on the temperature and should
be around 10%.
Segment WE-D (AC1MODE):
−
A symmetrical, gaussian-shaped peak should be plotted. The
evaluation must provide a result for the peak voltage U and the peak
current I, which are printed out at the end of the curve.
−
Peak voltage U :
– 450 mV ... – 550 mV