7 Safety, Errors, Troubleshooting, Diagnosis, GLP
746 VA Trace Analyzer / 747 VA Stand
7-50
7.6.2
Prepare instrument for diagnostic test
1 Power off.
2 Disconnect all cables at rear panel, except mains cable and connecting cable
to VA Stand.
3 Power on.
7.6.3
Checking thermal printer
1 Prepare instrument for diagnostic test (see section 7.6.2).
2 <VA TRACE ANALYZER>
3
<Diagn.>
4
<Thermoprinter>
5
<Test print>
A test printout is outputted (see following page).
•
Check printout for completeness and neatness.
•
The character set (the first five lines) must match the specimen.
•
The very top and very bottom points of a line must be printed clearly;
check K and p especially.
•
The printout must have a total length of 260 mm (tolerance:
±
10 mm)
6 Press <ESC> several times until the ”MONITORING” page appears.