7 Safety, Errors, Troubleshooting, Diagnosis, GLP

746 VA Trace Analyzer / 747 VA Stand

7-50

7.6.2

Prepare instrument for diagnostic test

1 Power off.

2 Disconnect all cables at rear panel, except mains cable and connecting cable

to VA Stand.

3 Power on.

7.6.3

Checking thermal printer

1 Prepare instrument for diagnostic test (see section 7.6.2).

2 <VA TRACE ANALYZER>

3

<Diagn.>

4

<Thermoprinter>

5

<Test print>

A test printout is outputted (see following page).

•

Check printout for completeness and neatness.

•

The character set (the first five lines) must match the specimen.

•

The very top and very bottom points of a line must be printed clearly;

check K and p especially.

•

The printout must have a total length of 260 mm (tolerance:

±

10 mm)

6 Press <ESC> several times until the ”MONITORING” page appears.