5.7 Program instructions
746 VA Trace Analyzer / 747 VA Stand
5-234
END
Page:
O, S
Stand:
M, A
Abbreviation:
E
Instructions t/s Main parameters Auxiliary parameters
––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––
## END .....
Function
End of the main program ("OP.SEQUENCE") or end of the segment
("SEGMENTS").
Description
The
END
instruction closes the main program on the "OP.SEQUENCE"
page or a segment on the "SEGMENTS" page.
Comments
•
Each program or segment must be closed with the
END
instruction.
•
The
END
instruction in the main program switches off all functions still
active at the 747 VA Stand (e.g. stirrer, purging, electrodes, voltage)
with exception of the standby voltage
U.standby
defined in a
ØMEAS
instruction.
•
In contrast to
END
in the main program, the
END
instruction in the
segment simply causes a return to the location in the main program
where the segment was called up.
see also
STIR, PURGE, DME, HMDE, SMDE, RDE, MEAS