5.7.3 Alphabetical instruction list
746 VA Trace Analyzer / 747 VA Stand
5-251
Description
The
PADD>M
instruction is used for the automatic or manual addition of
standard addition solutions. In contrast to the
ADD>M
instruction, the
standard addition volume
V.add
can not be preallocated, but is
recalculated in every measurement from the evaluation value
EV:<subst>
measured beforehand for any substance and a proportionality coefficient
Prop.coeff.
.
The automatic addition can be performed with a 685 Dosimat, a 700
Dosino or directly with the 695 Autosampler. The appropriate position of
the standard addition solution must be specified on the "SOLUTIONS"
page (see
Soln.name
)
Comments
•
The
PADD>M
instruction must be located within an ADD loop.
•
In the last execution of the ADD loop, the
PADD>M
instruction is no
longer executed.
•
On execution of the
PADD>M
instruction with a 685 Dosimat or 700
Dosino, a wait is made until the addition is complete before the
instruction following
PADD>M
is executed.