5.7.3 Alphabetical instruction list

746 VA Trace Analyzer / 747 VA Stand

5-227

DOS>F

Page:

O, S

Stand:

A

Abbreviation:

DOS>F

Instructions t/s Main parameters Auxiliary parameters

––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––

## .. DOS>F ..... Soln.name ........ V.add ........mL

Function

Dispense auxiliary solution into sample vessel (FEED) on the 695
Autosampler.

Parameter

Range; Default value / Meaning

Soln.name

8 characters; -

Name of the auxiliary solution

The entry of a solution name is mandatory when the auxiliary solution
should be added automatically. In this case the solution name

Soln.name

must be entered in the parameter of the same name on the

"SOLUTIONS" page and assigned a position

Pos.

(see section 5.3.4).

Without entry of a name for

Soln.name

in the

DOS>F

instruction, without

entry of this name on the "SOLUTIONS" page or without a position
assignment, the auxiliary solution must be dispensed manually (the
prompt

DOS "Soln.name" "V.add" > F

then appears during the

program run).

V.add

0.010...50.000 mL; 1.000 mL

Solution volume that should be dispensed

Volume of the auxiliary solution that should be dispensed into the
sample vessel.

Description

The

DOS>F

instruction is used for the manual or automatic addition of

auxiliary solutions (e.g. supporting electrolyte, reagent solution) to the
sample vessel (FEED) on the 695 Autosampler.

For automatic addition with the 695 Autosampler, the appropriate
position of the auxiliary solution must be specified on the "SOLUTIONS"
page (see

Soln.name

).

Comments

The

DOS>F

instruction can be used only in operation with the 695

Autosampler.

see also

DOS/M, DOS>M, DOS>A, PDOS>M, SDOS>M, SDOS>A, SDOS>F