Metrohm 746 VA Trace Analyzer User Manual

Page 144

Advertising
background image

4.5 Basic method for DP polarography

746 VA Trace Analyzer / 747 VA Stand

4-59

If you do not want an automatic printout in the following determination,

you can either delete this line with the softkey

<Delete line>

or clear

the output destination

Printer

for the line with the <CLEAR> key.

Press the softkey

<Comment>

. The subpage of "DOCUMENTATION"

appears with the method comment describing, among other things, the

procedure used in the determination.

Finally press the <ESC> key to return to the main page of "DOCUMEN-

TATION".

4.5.4

Performing the determination

To perform the determination, the following steps are necessary:

1

Switch to dialog page "MONITORING"

Press the yellow <MONITORING> key. The dialog page "MONITORING"

whose header shows the loaded method

POL

appears.

2

Switch the program run mode to

determination

If the parameter

Run mode

is set to

calibration

, proceed as follows:

Use the cursor keys to select the field

Run mode

.

Press the space bar to enter the edit mode.

Press the softkey

<determ.>

and confirm this entry with the <ENTER>

key.

3

Entry of the sample volume
Enter the sample volume as follows:

Use the cursor keys to select the first field below the title

Sample

size/S0

.

Press the space bar to enter the edit mode.

Enter

10 mL

and confirm this entry with the <ENTER> key.

4

Start the determination

Press the red <START> key.

The status display in the header of the dialog page "MONITORING"

switches to

* BUSY *

, the start time of the measurement is entered un-

der

Determination

and the determination is given a standard name,

also formed from the start time. Further, during the measurement the

measurement window is superimposed which shows the current pro-

gram line being executed and the remaining line time.

Shortly after the start of the determination, the screen shows the prompt

to add 10 mL sample to the measuring vessel.

5

Addition of sample

Remove stopper 25

25 on 747 VA Stand and use a pipette to add 10 mL of

drinking water to the measuring vessel through the pipetting opening.

Press the <ENTER> key.

The screen shows the prompt to add 1 mL acetate buffer to the measur-

ing vessel.