305 (var, O, s, M, a – Metrohm 746 VA Trace Analyzer User Manual
Page 456

5.7.3 Alphabetical instruction list
746 VA Trace Analyzer / 747 VA Stand
5-305
(VAR
Page:
O, S
Stand:
M, A
Abbreviation:
(V
Instructions t/s Main parameters Auxiliary parameters
––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––
## .. (VAR .....
Function
Start of the loop for parameter variation.
Description
The
(VAR
instruction opens a loop for the variation of a parameter
defined by the
VARNI
instruction. The instructions incorporated in the
VAR loop are executed repeatedly. The loop must be closed by the
VAR)n
instruction.
Comments
•
A program can contain several VAR loops, but they must not be
nested.
•
REP loops are allowed within VAR loops and a VAR loop can be
inserted in a REP loop.
•
No ADD or CAL loops are allowed within VAR loops.
•
Within VAR loops, the
EXIT
instruction for a premature exit from this
loop can be inserted.
Example
see
VARNI
see also
VAR)n, VARNI, (REP, REP)n