5.7.3 Alphabetical instruction list

746 VA Trace Analyzer / 747 VA Stand

5-311

VWASH

Page:

O, S

Stand:

A

Abbreviation:

VW

Instructions t/s Main parameters Auxiliary parameters

––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––

## .. VWASH ..... V.add ........mL

Function

Additional wash volume for subsequent dispensing or transfer instruction
on the 695 Autosampler.

Parameter

Range; Default value / Meaning

V.add

0.010...10.000 mL; 1.000 mL

Addition volume

Additional solution volume that is delivered following the dispensing or
transfer instruction.

Description

The

VWASH

instruction delivers an additional wash volume

V.add

of

hydraulic solution following a dispensing or transfer instruction on the
695 Autosampler. This ensures that the preceding solution transfer was
actually complete.

Comment

•

The

VWASH

instruction refers only to the immediately following

dispensing or transfer instruction.

•

The

VWASH

instruction can be used with the following dispensing or

transfer instructions for the 695 Autosampler:

SMPL>M
DOS>M, DOS>A, DOS>F, SDOS>M, SDOS>A, SDOS>F, PDOS>M
ADD>M, PADD>M, CSOLN>M, RADD>M, RADD>F
A>M, F>A, F>M

•

For solution transfers on the 695 Autosampler, the following generally
holds:

Transfer volume

VWASH

Siphoning off Delivery

< 200

µ

L

V.add

= 0

100%

100%

V.add

> 0

100%

100% +

V.add

≥

200

µ

L

V.add

= 0

105%

100% (+ 5% in waste)

V.add

> 0

100%

100% +

V.add

see also

DOS>A, DOS/M, DOS>M, DOS>F, PDOS>M, SDOS>M, SDOS>A,

SDOS>F, ADD/M, ADD>M, PADD>M, RADD>F, RADD>M

Advertising