3.10 Connection of 700 Dosinos

746 VA Trace Analyzer / 747 VA Stand

3-47

7 Connect siphoning Dosino to 747 VA Stand

Using either a 6.1805.120 FEP Tubing (L = 1 m) or two 6.1805.120 FEP

Tubings and a 6.1808.000 Connecting sleeve, connect connection 1 of

the 6.1567.250 Dosing unit of the siphoning Dosino with coupling 31

31

mounted on the 747 VA Stand (see Fig. 6).

Attach siphoning Dosino using the cable connected to the instrument to

socket 18

18 (Dos. 4) of the 746 VA Trace Analyzer (see Fig. 2 or Fig. 4).

8 Initialize Dosinos

Switch on 746 VA Trace Analyzer using the mains switch 11

11. The two

Dosing units of the rinsing and siphoning Dosino and the tubing connec-

tions are filled automatically at system start (the initialization procedure

can take several minutes!).