8 height/area calculation – Metrohm 746 VA Trace Analyzer User Manual

Page 482

Advertising
background image

6.8 Height/area calculation

746 VA Trace Analyzer / 747 VA Stand

6-19

6.8

Height/area calculation

With the aid of the calculated baselines, the evaluation quantity

EV:<subst>

is de-

termined for each substance and displayed as a result. Depending on the meas-

urement mode, the following quantities are possible as evaluation quantities:

Measurement

mode

I.wave

I.peak

P.peak

I.abs

I.rel

t.peak

DCT (waves)

n

DCT (peaks),
DP, SQW, AC1,

l

l

CYC (1 ramp)

AC2

l

l

PSA

n

l

adjustable

n

no other choice possible

These evaluation quantities are calculated as follows:

•

I.wave

The wave current

I.wave

is calculated

from the tangents before and after the

wave (details, see section 6.7.2).

I.wave

•

I.peak

The peak height

I.peak

is determined

from the previously calculated baseline

(details, see section 6.7.3) and the ex-

actly interpolated current height at the

peak maximum.

I.peak

•

P.peak

The peak area

P.peak

is determined

from the previously calculated baseline

(details, see section 6.7.3) by integra-

tion.
With a linear baseline with

Scope =

f.half

and

Scope = r.half

, only the

area up to the peak maximum is taken

as an evaluation quantity.

Scope = whole

Scope = f.half

Scope = r.half

•

I.abs / I.rel

With AC2 curves, either the absolute or the relative peak height can be cal-

culated. Details are given in section 6.7.4.

•

t.peak

With PSA curves, the peak area

t.peak

is determined by integration

of the previously calculated base-

line (details, see section 6.7.5).

P.peak

t.peak

Advertising