Metrohm 746 VA Trace Analyzer User Manual

Page 522

Advertising
background image

7.3 Error messages

746 VA Trace Analyzer / 747 VA Stand

7-23

Error message

Meaning/Causes

Rectification

Exit

SAMPLER - ERROR 1:
Missing device

Connection to 747 VA Stand or

695 Autosampler broken:

−

connecting cable not

plugged in

−

wrong address

−

hardware fault in VA Stand

Check connection to 747 VA Stand or

695 Autosampler:

−

plug in connecting cable

−

restart 746 VA Trace Analyzer

−

run diagnostic routine for VA Stand

E/S

SAMPLER - ERROR 2:
Hardware defect U1 or
U5 EBUS

EBUS interface faulty

Inform Metrohm service

E/C

SAMPLER - ERROR 3:
Transfer needle blocked

−

Transfer needle of 695

Autosampler blocked

−

695 Autosampler switched

off

−

Wrong response of RS232

interface of 695 Autosampler

−

Clean transfer needle

−

Switch on 695 Autosampler

−

Inform Metrohm service

E/S

SAMPLER - ERROR 4:
No reply from RS232

No response of RS232 inter-

face of 695 Autosampler

Inform Metrohm service

E/S

SAMPLER - ERROR 5:
RS232 command too long

Faulty RS232 transfer

Inform Metrohm service

E/S

SAMPLER - ERROR 6:
Command to RS232 not
sent

Faulty RS232 transfer

Inform Metrohm service

E/S

SAMPLER - ERROR 7:
RS232 communication
error

Faulty RS232 transfer

Inform Metrohm service

E/S

SAMPLER - ERROR 8:
Hardware defect U4 VART

Hardware fault in 695 Auto-

sampler

Inform Metrohm service

E/C

SAMPLER - ERROR 9:
No small syringe
detected

No electrical connection to

small syringe of 695 Autosam-

pler

Inform Metrohm service

E/C

SAMPLER - ERROR 10:
No large syringe
detected

No electrical connection to

large syringe of 695 Autosam-

pler

Inform Metrohm service

E/C

SAMPLER - ERROR 11:
No rack detected

−

No rack mounted on 695

Autosampler

−

No rack type defined for

mounted rack

−

Mount rack

−

Define rack type on the ”STAND/

Rack definition” page and store data

in 695 Autosampler

E/C

SAMPLER - ERROR 12:
Hardware defect U3
EEPROM

Hardware fault

Inform Metrohm service

E/C

SAMPLER - ERROR 13:
Hardware defect U9
D-Latch

Hardware fault

Inform Metrohm service

E/C

SAMPLER - ERROR 14:
Illegal position number

Inadmissible position number

for the mounted rack

Change position number

Pos.

on the

”SAMPLES” and ”MONITORING” pages

E/S

SAMPLER - ERROR 15:
Illegal rack number

The specified rack type number

is not admissible

Define rack type on the ”STAND/Rack

definition” page and store data in 695

Autosampler

E/S