302 tstir, O, s, M, a – Metrohm 746 VA Trace Analyzer User Manual
Page 453

5.7 Program instructions
746 VA Trace Analyzer / 747 VA Stand
5-302
TSTIR
Page:
O, S
Stand:
M, A
Abbreviation:
TS
Instructions t/s Main parameters Auxiliary parameters
––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––
## .. TSTIR ..... Rot.speed ......../min
Function
Switch on stirrer or a preset line time.
Parameter
Range; Default value / Meaning
t/s
0.1...999.9 s; -
Stirring time
The duration of stirring is entered in the line time.
Rot.speed
0, 200, 400, ..., 3000 /min; 2000 /min
Rotational speed
Setting of the number of revolutions of the stirrer per min.
Description
The
TSTIR
instruction is used to switch on the stirrer of the 747 VA Stand
for the period preset in
t/s
.
Comment
If the stirrer should not be switched off automatically after a definite time,
the
STIR
instruction can also be used.
see also
STIR, ØSTIR