5 Operation
746 VA Trace Analyzer / 747 VA Stand
5-8
Key
Meaning
ABBREV
Abbreviate current instruction
The line time of the current program instruction is set to 0 s and the next
instruction is started.
Exception: With the SWEEP instructions the line time is also
set to 0, but the sweep is run to completion.
HOLD
(STEP)
Hold = Program interruption
The current program instruction is stopped. The display
* BUSY (HOLD)
*
then appears on the "MONITORING" page instead of
* BUSY *
. The
program can be continued by pressing the <(CONT)> key at the
location where it was interrupted.
(Step) = Skip to next instruction
The program instruction interrupted with <HOLD> is run to completion
and a skip then made to the next instruction. This instruction is exe-
cuted (without processing of the line time) and the program is then
interrupted again.
ENTER
Enter key
This key is used to close every field entry and confirm the current
parameter value.