Metrohm 746 VA Trace Analyzer User Manual
Page 270

5.5.3 CURVES
746 VA Trace Analyzer / 747 VA Stand
5-119
Cyclic Segment-VR-Curve: XXXXXXXX-XX-original
Modify
column
Start
zoom
New
display
Reset
param.
Save
param.
Softkeys
Meaning
Modify
column
Modify display/plot parameters column-by-column
Modification of the parameters for the axis scaling of the selected
cyclic segment. For details of the display/plot parameters, see section
5.6.2.
Modifications made here initiate a recalculation and curve display on
closure of the entry. These changes are temporary, they are canceled
with
<Reset param.>
or after exit with <ESC>.
The changes are permanently incorporated in the method only if they
are stored with
<Save param.>
before exiting.
Start
zoom
Select zoom range
This softkey starts the selection of the zoom region (curve section
which should be enlarged).
A segment curve can be zoomed only in the y direction,
the x axis is always shown full scale (limit values
U.start
and
U.end
of the segment).
First, the cursor keys <è>, <ç>, <é> and <ê> must be used
to define the top left corner point of the zoom region (marked by a
cross) as follows:
Each time these keys are pressed, the corner point moves in the
selected direction by 5 pixels. A finer shift of 1 pixel can be
achieved when the <SHIFT> key is pressed at the same time. In
both cases, the cursor keys are repetitive, i.e. the longer the cur-
sor key is pressed, the further the corner point is shifted.
16
Display / Plot: I = f(U)
––––––––––––––––––––––––––
I.scale ........
I.begin ........A
I.end ........A
U.div ........V/cm
15
ESC