Metrohm 746 VA Trace Analyzer User Manual

Page 445

Advertising
background image

5.7 Program instructions

746 VA Trace Analyzer / 747 VA Stand

5-294

The phase dependent differences in the current measurements give
peak-shaped curves which can be evaluated using the baselines of type

linear

,

polynom

or

expon.

(see section 5.6.3).

I

U

U.peak

I.peak

Comments

•

The

SQWMODE

measurement mode can be used with all types of

electrodes together with the instructions

MEAS

,

SWEEP

,

DSWEEP

,

FSWEEP

or

TSWEEP

.

•

The electrode type

DME

,

SMDE

,

HMDE

or

RDE

must be entered before

the

SQWMODE

measurement mode.

•

If the

SQWMODE

measurement mode is used with the voltage sweep

SWEEP

and the electrode type

HMDE

, the potentiostat normally oper-

ates in the operating mode "slow". A switch is made to the operating
mode "fast" when the modulation frequency

Modul.freq.

>120 Hz.

•

If the

SQWMODE

measurement mode is used with the voltage sweep

SWEEP

and the electrode type

SMDE

or

DME

, the setting of the poten-

tiostat depends on the measurement profile. Normally, a switch is
also made from "slow" to "fast" when the modulation frequency

Modul.freq.

>120 Hz. With very small signals, however, the operat-

ing mode "slow" is used and with very large signals the operating
mode "fast", irrespective of the pulse time.

•

If the

SQWMODE

measurement mode is used with the voltage sweep

SWEEP

and the electrode type

RDE

, the potentiostat always operates in

the operating mode "fast".

•

If the

SQWMODE

measurement mode is used with the voltage sweep

FSWEEP

, the potentiostat always operates in the operating mode "fast"

with all electrode types.

•

The quantity

t.drop

important for the definition of

t.step

depends

on the type of electrode selected:

DME

t.drop = 20 ms

HMDE and RDE

t.drop = 0 ms

SMDE

t.drop = Drop.size

×× 40 ms

see also

SWEEP, FSWEEP, DME, SMDE, HMDE, RDE

n = Meas.cycles

Σ

(I

n1

– I

n2

)

n

I

=