3.4 Multi-mode electrode (MME)

746 VA Trace Analyzer / 747 VA Stand

3-23

3.4.7

Replenishing the mercury (without changing capillary)

The multi-mode electrode 38

38 can also be refilled with mercury without having to

remove the glass capillary 109

109.

1 Dismantle multi-mode electrode

Unscrew FEP tubing 51

51 and 59

59 from the MME. Disconnect electrode

cable 37

37 from the MME.

Take multi-mode electrode 38

38 out of the measuring head 41

41 and tap the

MME gently to knock off any mercury drops on the glass capillary into the

measuring vessel.

Place multi-mode electrode 38

38 horizontally in the electrode holder 114

114

(see Fig. 15-3). The slotted screw 101

101 is now at the top.

2 Replenish mercury

Unscrew slotted screw 101

101 using a suitable coin. If the slotted screw 101

101

can not be loosened by hand, screw on locking ring 95

95 and pull out of the

MME.

Draw up mercury using the syringe 111

111 with attached needle 112

112 and

expel into the Hg reservoir 103

103.

The Hg reservoir 103

103 must never be filled more than

2

/

3

full with

mercury.

Reinsert slotted screw 101

101 into screw thread 100

100 and screw flush to

surface using a suitable coin (this action may expel a few drops of mer-

cury from glass capillary 109

109).

Do not turn so tightly that the cemented-in steel threaded ring 100

100

becomes loose and hence jeopardizes the tightness and safety of the

MME!