7.7 Standard operating procedure in the context of GLP/ISO 9001 guidelines

746 VA Trace Analyzer / 747 VA Stand

7-61

7.7.4

Performing the determination

To perform the lead determination in the prepared sample solution, the following

steps are necessary:

1 Switch to dialog page ”MONITORING”

Press the yellow <MONITORING> key.

2 Switching the program run mode to

determination

If the paramete

r Run mode

is set to

calibration

, proceed as follows

Use the cursor keys to select the field

Run mode

.

Press the <CLEAR> key.

Press the softkey

<determ.>

and confirm this entry with the <ENTER>

key.

3 Entry of the sample volume

Enter the sample volume as follows:

Use the cursor keys to select the first field below the title

Sample size/S0

.

Press the <CLEAR> key.

Enter

0.1 mL

and confirm this entry with the <ENTER> key.

4 Start the determination

Press the red <START> key.

The sample solution is measured three times. On completion of the third

measurement, the program is interrupted. The screen shows the prompt

to add 50

µ

L of the standard addition solution

PbStd

to the measuring

vessel

ADD PbStd 50.000 uL > M

.

5 First standard addition

Remove stopper 25

25 and add 50

µ

L of the lead ion standard solution to

the measuring vessel through the pipetting opening using a pipette. Then

close the pipetting opening with stopper 25

25 and press the <ENTER>

key.

The spiked sample solution is measured three times. On completion of

the third measurement, the program is interrupted. The screen again

shows the prompt to add 50

µ

L of standard addition solution

PbStd

to

the measuring vessel

ADD PbStd 50.000 uL > M

.

6 Second standard addition

Remove stopper 25

25 and add 50

µ

L of the lead ion standard solution to

the measuring vessel through the pipetting opening using a pipette. Then

close the pipetting opening with stopper 25

25 and press the <ENTER>

key.

The sample solution spiked a second time is measured three times. On

completion of the third measurement, the results are calculated and the

reports and curves printed out (see section 4.2).