Metrohm 746 VA Trace Analyzer User Manual
Page 313

5.6 Method Specifications
746 VA Trace Analyzer / 747 VA Stand
5-162
Parameter
Range; Default value / Meaning
Instrument variables
The following instrument variables can be used in formulae:
C##
Common variable C01...C18 from the working
memory (see section 5.4.5). The common vari-
ables can be used to exchange results between
different determinations.
ED:<subst>:<V>
Difference of the mean values of the evaluation
quantities (
I.delta
,
P.delta
or
t.delta
) for the
specified substance
<subst>
and variation
<V>
.
The numeric value put out refers to the basic
unit A, W or s.
ES:<subst>:<V>
Standard deviation of evaluation quantities
(
Std.dev.
) for the specified substance
<subst>
and variation
<V>
. The numeric value put out
refers to the basic unit A, W or s.
EU:<subst>:<V>
Mean value of peak voltages (
U
) or half-wave
voltages (
U/2
) for the specified substance
<subst>
and variation
<V>
. The numeric value
put out refers to the basic unit V.
EV:<subst>:<V>
Mean value of evaluation quantities (
I.mean
,
P.mean
or
t.mean
) for the specified substance
<subst>
and variation
<V>
. The numeric value
put out refers to the basic unit A, W or s.
MA:<subst>
Standard addition mass (
Add.mass
) for the
specified substance
<subst>
. The numeric
value put out refers to the basic unit g.
MC:<subst>
Mass concentration
ρ
s
(
Mass conc.
) for the
specified substance
<subst>
. The numeric
value put out refers to the basic unit g/L.
MD:<subst>
Total scatter (
MC.dev.
) of mass concentration
ρ
s
for the specified substance
<subst>
. The nu-
meric value put out refers to the basic unit g/L.
ML:<subst>
Scatter due to calibration (
Cal.dev.
) of mass
concentration
ρ
s
for the specified substance
<subst>
. The numeric value put out refers to the
basic unit g/L.
MS:<subst>
Substance mass (
Mass
) in the measuring vessel
for the specified substance
<subst>
. The nu-
meric value put out refers to the basic unit g.
KD:<subst>
Mean scatter (
Mean deviat.
) for calibration and
standard addition curve of the specified sub-
stance
<subst>
. The numeric value put out re-
fers to the basic unit A, W or s.
KN:<subst>
Non-linearity factor (
Nonlin.
) for calibration and
standard addition curve of the specified sub-
stance
<subst>
. The numeric value put out re-
fers to the basic unit [AL/g]
4
, [WL/g]
4
or
[sL/g]
4
.