Metrohm 746 VA Trace Analyzer User Manual
Page 384

5.7.3 Alphabetical instruction list
746 VA Trace Analyzer / 747 VA Stand
5-233
Description
The
DSWEEP
instruction (Dummy Sweep) starts the voltage sweep
defined in the parameters. While measured values are recorded during
the sweep, they are not stored.
U
t
U.start
U.end
U.step
t.step
Comments
•
The electrode type must be defined before the
DSWEEP
instruction.
Admissible are
DME
,
SMDE
,
HMDE
or
RDE
.
•
The measurement mode must be defined before the
DSWEEP
instruction. Admissible are
DCTMODE
,
DPMODE
,
AC1MODE
,
AC2MODE
or
SQWMODE
, but not
PSAMODE
or
CYCMODE
.
•
With the
DSWEEP
instruction, the potentiostat always operates in the
operating mode "fast".
•
If the final voltage
U.end
should not be switched off at the end of the
program, enter a suitable standby voltage
U.standby
in a
ØMEAS
instruction.
see also
SWEEP, FSWEEP, SSWEEP, DME, SMDE, HMDE, RDE, DCTMODE,
DPMODE, AC1MODE, AC2MODE, SQWMODE