5.7 Program instructions
746 VA Trace Analyzer / 747 VA Stand
5-238
F>M
Page:
O, S
Stand:
A
Abbreviation:
F>M
Instructions t/s Main parameters Auxiliary parameters
––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––
## .. F>M ..... addition FEED -> MEAS V.add ........mL
Function
Transfer a solution from the sample vessel (FEED) on the 695 Autosam-
pler to the measuring vessel (MEAS) on the 747 VA Stand.
Parameter
Range; Default value / Meaning
V.add
0.010...50.000 mL; 1.000 mL
Transfer volume
Solution volume that should be transferred from the sample vessel on
the sample rack of the 695 Autosampler to the measuring vessel on the
747 VA Stand.
Description
The
F>M
instruction transfers the solution volume specified under
V.add
from the sample vessel (FEED) on the 695 Autosampler to the measur-
ing vessel (MEAS) on the 747 VA Stand.
Comments
•
The
F>M
instruction can be used only in operation with the 695
Autosampler.
•
If the solution volume in the sample vessel <
V.add
, an error
message appears.
see also
A>M
,
F>A