background image

4.4 Lead determination with internal standard

746 VA Trace Analyzer / 747 VA Stand

4-47

Report Full

Printout of the full result report on the printer

Curve Cal Sbst:Pb

Printout of the calibration curve on the printer

Press the softkey

<Comment>

. The subpage of "DOCUMENTATION"

appears with the method comment describing, among other things, the

procedure used in the calibration method with the internal standard.

Finally press the <ESC> key to return to the main page of "DOCUMEN-

TATION".

4.4.4

Recording the calibration curve

To be able to perform a determination with the internal standard, the proportionality

factor must first be determined using a calibration curve. The following steps are

necessary for this:

1

Switch to dialog page "MONITORING"

Press the yellow <MONITORING> key. The dialog page "MONITORING"

whose header shows the loaded method

INTSTD

appears.

2

Switch the program run mode to

calibration

If the parameter

Run mode

is set to

determination

, proceed as follows:

Use the cursor keys to select the field

Run mode

.

Press the space bar to enter the edit mode.

Press the softkey

<calibr.>

and confirm this entry with the <ENTER>

key.

3

Entry of the sample volume
Enter the sample volume as follows:

Use the cursor keys to select the first field below the title

Sample

size/S0

.

Press the space bar to enter the edit mode.

Enter either

*

or

10 mL

(as preparation for performance of the determi-

nation in section 4.4.5) and confirm this entry with the <ENTER> key.

Although the sample volume for recording the calibration curve is not needed (the
volume of the calibration solutions is defined in the parameter
V0.CSoln), the
Sample size/S0 field must not be empty.

4

Start the determination

Press the red <START> key.

The status display in the header of the dialog page "MONITORING"

switches to

* BUSY *

, the start time of the measurement is entered un-

der

Determination

and the determination is given a standard name,

also formed from the start time. Further, during the measurement the

measurement window is superimposed which shows the current pro-

gram line being executed and the remaining line time.