4 Introduction to the operation

746 VA Trace Analyzer / 747 VA Stand

4-64

5

Switch to dialog page "DOCUMENTATION"

•

Press the yellow <DOCUMENTATION> key. The dialog page "DOCU-

MENTATION" appears for the automatic output of reports and curves.

The line entered here has the following meaning:

Curve Smth Sgmt:ASV VR:00

Printout of the smoothed segment curve
00 on the printer

•

If you do not want an automatic printout in the following determination,

you can either delete this line with the softkey

<Delete line>

or clear

the output destination

Printer

for the line with the <CLEAR> key.

•

Press the softkey

<Comment>

. The subpage of "DOCUMENTATION"

appears with the method comment describing, among other things, the

procedure used in the determination.

•

Finally press the <ESC> key to return to the main page of "DOCUMEN-

TATION".

4.6.4

Performing the determination

To perform the determination, the following steps are necessary:

1

Switch to dialog page "MONITORING"

•

Press the yellow <MONITORING> key. The dialog page "MONITORING"

whose header shows the loaded method

ASV

appears.

2

Switch the program run mode to

determination

If the parameter

Run mode

is set to

calibration

, proceed as follows:

•

Use the cursor keys to select the field

Run mode

.

•

Press the space bar to enter the edit mode.

•

Press the softkey

<determ.>

and confirm this entry with the <ENTER>

key.

3

Entry of the sample volume
Enter the sample volume as follows:

•

Use the cursor keys to select the first field below the title

Sample

size/S0

.

•

Press the space bar to enter the edit mode.

•

Enter

10 mL

and confirm this entry with the <ENTER> key.