5.7 Program instructions

746 VA Trace Analyzer / 747 VA Stand

5-264

PURGE

Page:

O, S

Stand:

M, A

Abbreviation:

PU

Instructions t/s Main parameters Auxiliary parameters

––––––––––––– ––––– ––––––––––––––––––––––––– –––––––––––––––––––––––––

## .. PURGE .....

Function

Switch on purging.

Description

The

PURGE

instruction switches on the inert gas purging for the

measuring vessel on the 747 VA Stand. This is used to remove the
electrochemically active and hence interfering oxygen in aqueous
solutions. With the inert gas flow rate of ca. 20 L/h set on the 747 VA
Stand, a purging time of ca. 3...5 min usually suffices.

Purging remains active until it is switched off automatically with the

ØPURGE

or

END

instruction in the main program on the "OP.SEQUENCE"

page.

Comment

If the purging should be switched off automatically after a definite
time, the

TPURGE

instruction can also be used.

To achieve an effective purging of the analysis solution as quickly as
possible, the solution should also be stirred with the

STIR

or

TSTIR

instruction.

see also

ØPURGE, TPURGE, STIR, TSTIR, END