6.9 Content calculation

746 VA Trace Analyzer / 747 VA Stand

6-27

•

Working in the nonlinear range

If the substance content is in the nonlinear range, several standard additions

have to be performed to determine the nonlinear curve shape exactly.

•

Standard addition ratio 1:2 to 1:5

The optimum standard addition ratio for the entire standard addition is from

1:2 to 1:5, i.e. the sum of all standard addition amounts should be 2 to 5

times the amount of sample in the measuring vessel. This can be easily

checked later with the parameters

Mass

and

Add.mass

put out on the dialog

page "RESULTS/Evaluations".

•

Consideration of blank values

Any blank values must be determined separately and subtracted in the for-

mula calculation on the dialog page "CALCULATIONS".

6.9.5

Calibration curve

The content determination using a calibration curve is performed in two steps:

– First the relation between the mass concentration

ρ

eff

eff

of a sub-

stance and the evaluation quantity

EV

is determined by measur-

ing different reference solutions.

– Finally, the sample is measured and its mass concentration

ρ

eff,s

eff,s

determined using the recorded calibration curve.

The recording of a calibration curve is programmed with the following instructions

and parameters on the "OP.SEQUENCE" page (see section 5.7.3):

(CAL...CAL)n

Loop for recording the calibration curve

n

Number of calibration solutions – 1

CSOLN...

Addition of the calibration solutions

The remaining parameters needed for the calibration curve are entered on the dia-

log page "SUBSTANCES/Calibration" (see section 5.6.3). The most important of

these are:

Technique

Selection of the calibration technique

cal.crv.

Calibration curve

Curve type

Curve type

linear

Straight line of type

y = a + bx

0/linear

Straight line of type

y = bx

nonlin

Curve of type

y = a + bx + cx

4

0/nonlin

Curve of type

y = bx + cx

4

V0.CSoln.

Addition volume of the calibration solutions

CSoln.name

Name of the calibration solutions

Mass conc.

Mass concentration of the calibration solutions

In the content determination using a calibration curve, the following procedure is

used to calculate the sought mass concentration

ρ

eff,s

eff,s

of the sample (see also ex-

ample in section 4.3):