4.3 Lead determination with calibration curve

746 VA Trace Analyzer / 747 VA Stand

4-31

4.3.3

Method specifications of the method

CALCRV.mth

In this section the 5 dialog pages of the group "Method Specifications" associated

with the loaded method

CALCRV.mth

will be explained in detail. If you are interested

only in a rapid lead measurement, you can proceed directly to section 4.3.4.

1

Switch to dialog page "OPERATION SEQUENCE"

•

Press the yellow <OP. SEQUENCE> key. The dialog page "OPERATION

SEQUENCE" appears with the main program for the lead determination:

Method: CALCRV .mth OPERATION SEQUENCE

Title : Pb determination with calibration curve

Instructions t/s Main parameters Auxiliary parameters

ДДДДДДДДДДДДД

ДДДДДДДДДДДДД ДДДДД

ДДДДД ДДДДДДДДДДДДДДДДДДДДДДДДД

ДДДДДДДДДДДДДДДДДДДДДДДДД

ДДДДДДДДДДДДДДДДДДДДДДДДД

ДДДДДДДДДДДДДДДДДДДДДДДДД

1 D SMPL>M V.fraction mL V.total L

2 C (CAL

3 C CSOLN>M addition into MEAS

4 DOS>M Soln.name Acetate V.add 1.000 mL

5 PURGE

6 STIR 180.0 Rot.speed 2000 /min

7 ØPURGE

8 (REP

9 STIR 5.0 Rot.speed 2000 /min

10 SEGMENT Segm.name DPASV

11 REP)2

12 C RINSE Cycles 1 V.rinse mL

13 C CAL)3

14 END

Modify

column

Modify

line

Delete

instr.

Insert

instr.

Method

labels

•

The following is a brief explanation of the main program sequence:

Line 1

Addition of sample: 10 mL

Line 2

…

…13

Loop for recording the calibration curve (4 calibration solutions)

Line 3

Addition of calibration solution to measuring vessel: 10 mL

Line 4

Addition of acetate buffer to measuring vessel: 1 mL

Line 5

…

… 6

Purging and stirring of the analysis solution for 3 min

Line 7

Switching off the purging

Line 8

…

…11

Loop for double replication of the segment

Line 9

Stirring of the analysis solution for 5 s

Line 10

Call-up of the segment

DPASV for lead determination

Line 12

Resetting of the volume bookkeeping to 0 for solution change
(new calibration solution)

Line 14

End of main program

•

The sequence of the main program depends on the selected mode for

the program run

Run mode

on the "MONITORING" page:

−

In the program run mode

calibration

, all instructions marked with

D

are ignored. 4 calibration solutions are measured in turn and the

coefficients of the calibration curve determined.

−

In the program run mode

determination

, on the other hand, all in-

structions marked with

C

are ignored. The sample is measured and

its mass concentration then determined using the calibration curve

recorded earlier.