7.3 Error messages

746 VA Trace Analyzer / 747 VA Stand

7-25

Error message

Meaning/Causes

Rectification

Exit

SEGM.<segm> - ERROR 5:
Instruction not allowed
in block: line ##

There is an inadmissible

instruction in the specified

program line

Delete inadmissible instruction in the

segment

<segm>

Esc

SEGM.<segm> - ERROR 6:
Missing part of prep.
(PREP, PREP), >PREP

One of the three instructions for

a PREP block missing

In program in the segment

<segm>

,

check that all three instructions

(PREP

,

PREP)

and

>PREP

are present

Esc

SEGM.<segm> - ERROR 8:
Invalid with VARNI:
line ##

Inadmissible parameter for

parameter variation

Check parameter name for

Param.name

in the segment

<segm>

and correct

Esc

SEGM.<segm> - ERROR 9:
Invalid with (CSWEEP:
line ##

Inadmissible instruction in the

CSWEEP block (only

RAMP

is

allowed)

Delete inadmissible instruction in the

CSWEEP block in the segment

<segm>

Esc

SEGM.<segm> - ERROR 10:
Inconsist. value in
.....: line ##

Inadmissible value for the

specified parameter

Change parameter value in the segment

<segm>

Esc

SEGM.<segm> - ERROR 11:
Value inconsist. with
electrode: t.step in
line ##

t.step

value for specified

electrode too low

Increase

t.step

or reduce

t.modul

Change electrode type

Esc

SEGM.<segm> - ERROR 12:
Missing value in line
##

Missing parameter value in the

specified program line

Entry parameter value in the segment

<segm>

Esc

SEGM.<segm> - ERROR 13:
Not allowed data-
generating: line ##

Inadmissible data-generating

instruction in the specified

program line

Delete inadmissible instruction in the

segment

<segm>

Esc

SEGM.<segm> - ERROR 14:
Not allowed in segment:
line ##

Inadmissible instruction in the

segment

Delete inadmissible instruction in the

segment

<segm>

Esc

SEGM.<segm> - ERROR 16:
Incorrect use of (CAL
... RINSE sequence

If several CAL loops are used in

a program, every loop must

have a

RINSE

instruction

Enter

RINSE

instructions in the seg-

ment

<segm>

Esc

SEGM.<segm> - ERROR 17:
Sweep range too small

Sweep range too small

(< 8 measured values)

Increase sweep range or decrease

U.step

or

t.step

Esc

SEGM.<segm> - ERROR 18:
Sweep range too large

Sweep range too large

(> 1000 measured values)

Decrease sweep range or increase

U.step

or

t.step

Esc

SEGM.<segm> - ERROR 19:
Missing t.step

t.step

value missing or is

<1 ms (possible only in

variation loops)

Enter variation loop for

t.step

in the

segment

<segm>

so that

t.step

remains >1 ms

Esc

SEGM.<segm> - ERROR 20:
Conditioning cycle
range too small

Initial voltage

U.start

and

final voltage

U.end

of the

CONDC

instruction are identical

Change

U.start

and/or

U.end

in the

CONDC

instruction

Esc

SOLUTIONS - ERROR 1:
Mass conc. conflict on
line ##

The mass concentrations of

solutions with the same name

do not match

Check mass concentrations

Mass

conc.

of the solutions on the

”SOLUTIONS”, ”OP.SEQUENCE”,

”SEGMENTS” and ”SUBSTANCES/

Calibration” page and change

Esc

STORAGE - ERROR 1:
Destination file
secured

Secured file can not be deleted

or overwritten

Unsecure file before deleting or over-

writing with

<Secure/Unsecure>

E/C

STORAGE - ERROR 2:
Can not find source

The file to be copied is not

available at the source speci-

fied

Check source (working storage or

RS232 interfaces) to see whether the file

to be copied exists

E/C

STORAGE - ERROR 3:
Entry table full

Maximum number of files (256)

reached

Delete old files

E/C